VitaGuard® VG 3100 - Heart, Apnea and SpO2 MonitorGeneral
| Weight with power pack |
approx. 700 g |
| Dimensions |
13,5 x 20,3 x 4,5 cm |
| Casing material |
ABS |
| Power supply |
Power pack (4,8V, NiMH) or |
|
4 x batteries (1,5V, LR6, alkaline) or |
|
9 V power adapter NA 3000-2 |
| Power adapter NA 3000-2 |
Input: 100–240V, 50–60 Hz, 400 mA |
|
Output: 9 V d. c., 1,5 A |
| Power pack recharge time |
< 6 h |
| Operating time |
> 8 hours with alkaline batteries or NiMH power pack |
| Replace battery message |
Message on LCD display |
| Battery exhaustion message |
Visual and acoustic warning |
| Keys |
6 pushbuttons |
| ECG / respiration connector |
8-pin miniature round connector, type CF input |
| SpO2 connector |
14-pin mini-ribbon connector, type BF input |
| USB connector |
Mini USB to connect with a PC |
| AUX connector features |
RS232 interface for modem or online output |
| External alarm unit output
| | Nurse call unit output | |
2 analog inputs (AUX 1 & 2) |
| Display elements |
5 LED's and a graphical LCD display (320 x 240 dots) with | | backlight when powered by external or automobile power adapter |
| Alarm warnings |
Visual and acoustic as per DIN EN 60601-2-49 | | and IEC 60601-1-8 |
Apnea Monitor
| Method |
impedance pneumography |
| Respiration range |
4 – 120 breaths / min,
depending on age group selected |
| Resolution |
1 / min |
| Apnea pause settings |
8, 10, 12 ... 30, 32, 34 s |
| End of alarm condition | 2 breaths within 6 s |
| Minimum amplitude |
approx. 0,2 ohm |
| Signal recognition |
Green LED and selectable beep tone |
Heart Rate Monitor
| Heart rate range |
20–270 / min |
| Resolution |
1 / min |
| Bradycardia alarm settings |
30, 35 ... 175, 180 / min |
| Tachycardia alarm settings |
100, 105 ... 250, 255 / min |
| Minimum amplitude |
approx. 0,2 mV | | Selectable ECG leads | Einthoven I, II or III |
| Signal recognition |
green LED and selectable beep tone |
Pulse Rate Monitor
| Method |
pulse oximetry |
| Pulse rate range |
25 –240 / min |
| Resolution |
1 / min |
| Accuracy |
± 3 digits without motion
|
| ± 5 digits during motion
|
| Alarm settings |
same as for heart rate monitor |
SpO2 Monitor
| SpO2 range |
1–100 % |
| Resolution |
1 % |
| Accuracy |
±3 digits for SpO2 above 70 %
|
| Alarm settings |
Selectable from 50-100 % |
| Sensitivity |
Minimum (APOD), Standard, Maximum | |
(APOD =Adaptive Probe Off Detection) |
Memory
| Storage functions |
Event, trend, full-disclosure and compliance memory
|
| Storage capacity |
approx. 200 events, depending on pre-alarm and post-alarm settings, |
|
72 h trend, 8 h full-disclosure |
| Stored data |
ECG, heart rate, respiration, basal impedance, SpO22, |
|
pulse rate, signal IQ, plethysmogram, perfusion index, status, 2 analog inputs |
Classifications
| Product classifications |
IIb as per MDD 93/42/EEC
|
| Electrical classification |
Class II as per DIN EN 60601-1 |
| EMC classification |
Group 1, class B as per DIN EN 60601-1-2
|
| Ingress protection |
IP 21 as per DIN EN 60529 |
| German Hilfsmittelnummer |
21.24.02.5005 (applies to Germany only)
|
Environmental Conditions
| Operating temperature range |
5–40 °C |
| Operating humidity range |
5-95 %, non-condensing |
| Altitude |
- 300 m to 5000 m |
| Storage temperature range |
- 40 to 70 °C |
Subject to change.
|