DescriptionAnalysis SoftwareTechnical DataProduct Inquiry
Download
Products
Home
Site map
Imprint
Login: 

Password: 

VitaGuard® VG 3100 - Heart, Apnea and SpO2 Monitor

General

Weight with power pack approx. 700 g
Dimensions 13,5 x 20,3 x 4,5 cm
Casing material ABS
Power supply Power pack (4,8V, NiMH) or
4 x batteries (1,5V, LR6, alkaline) or
9 V power adapter NA 3000-2
Power adapter NA 3000-2 Input: 100–240V, 50–60 Hz, 400 mA
Output: 9 V d. c., 1,5 A
Power pack recharge time < 6 h
Operating time > 8 hours with alkaline batteries or NiMH power pack
Replace battery message Message on LCD display
Battery exhaustion message Visual and acoustic warning
Keys 6 pushbuttons
ECG / respiration connector 8-pin miniature round connector, type CF input
SpO2 connector 14-pin mini-ribbon connector, type BF input
USB connector Mini USB to connect with a PC
AUX connector features RS232 interface for modem or online output
External alarm unit output
Nurse call unit output
2 analog inputs (AUX 1 & 2)
Display elements 5 LED's and a graphical LCD display (320 x 240 dots) with
backlight when powered by external or automobile power adapter
Alarm warnings Visual and acoustic as per DIN EN 60601-2-49
and IEC 60601-1-8

Apnea Monitor

Method impedance pneumography
Respiration range 4 – 120 breaths / min,
depending on age group selected
Resolution 1 / min
Apnea pause settings 8, 10, 12 ... 30, 32, 34 s
End of alarm condition2 breaths within 6 s
Minimum amplitude approx. 0,2 ohm
Signal recognition Green LED and selectable beep tone

Heart Rate Monitor

Heart rate range 20–270 / min
Resolution 1 / min
Bradycardia alarm settings 30, 35 ... 175, 180 / min
Tachycardia alarm settings 100, 105 ... 250, 255 / min
Minimum amplitude approx. 0,2 mV
Selectable ECG leadsEinthoven I, II or III
Signal recognition green LED and selectable beep tone

Pulse Rate Monitor

Method pulse oximetry
Pulse rate range 25 –240 / min
Resolution 1 / min
Accuracy ± 3 digits without motion
± 5 digits during motion
Alarm settings same as for heart rate monitor

SpO2 Monitor

SpO2 range 1–100 %
Resolution 1 %
Accuracy ±3 digits for SpO2 above 70 %
Alarm settings Selectable from 50-100 %
Sensitivity Minimum (APOD), Standard, Maximum
(APOD =Adaptive Probe Off Detection)


Memory

Storage functions Event, trend, full-disclosure and compliance memory
Storage capacity approx. 200 events, depending on pre-alarm and post-alarm settings,
72 h trend, 8 h full-disclosure
Stored data ECG, heart rate, respiration, basal impedance, SpO22,
pulse rate, signal IQ, plethysmogram, perfusion index, status, 2 analog inputs

Classifications

Product classifications IIb as per MDD 93/42/EEC
Electrical classification Class II as per DIN EN 60601-1
EMC classification Group 1, class B as per DIN EN 60601-1-2
Ingress protection IP 21 as per DIN EN 60529
German Hilfsmittelnummer 21.24.02.5005 (applies to Germany only)

Environmental Conditions

Operating temperature range 5–40 °C
Operating humidity range 5-95 %, non-condensing
Altitude - 300 m to 5000 m
Storage temperature range - 40 to 70 °C

Subject to change.




Medizin- und Informationstechnik AG
Oderstrasse 77
14513 Teltow
Germany

Telephone +49 3328 3942-0
Telefax +49 3328 3942-99
E-Mail info@getemed.de